Biowave 3+ CT Spectrophotometer

biowave 3+ CT color touch spectrophotometer biochrom

Biowave 3+ CT Spectrophotometer

Product name
Item #

 

The Biowave 3+ Color Touch is truly an all-encompassing Life Sciences Spectrophotometer.

From the most basic cell density or single wavelength reading, to low volume molecular quantification, this is the most versatile single beam UV/Visible instrument on the market.

This compact and robust instrument is perfect for most Life Science laboratories, from undergraduate teaching to biotech and R&D, the Biowave 3+ CT can cover the full spectrum of needs.

 

DOWNLOAD BROCHURESCHEDULE A MEETING

Product name
Item #
biochrom
Product name
Item #

 

 

Built in Life Science methods

Pre-programmed methodologies for nucleic acid quantification (DNA, RNA and oligonucleotides), protein assays (BCA, Biuret, Bradford and Lowry) and for cell culture density (OD600) measurements. The visualization of the nucleic acid scan is particularly useful, especially for RNA samples where impurities may be present in the 230 nm region, yet not have an adverse effect on the A260/A280 ratio. The system is compatible with low volume UV cuvettes. For those samples that need the most precise results, the Biowave 3+ CT also comes loaded with a menu of commonly available fluorescent dyes to specifically target nucleic acids and proteins.

In kinetics mode, the basic plot of absorbance against time may be supplemented with the result for A/min, plus the correlation coefficient is also calculated for the duration of the assay. This slope may be multiplied automatically by a factor to convert it directly to rate of reaction.

Product name
Item #

Flexible Data Viewing and Export Options


The 5 inch color touch screen allows users to distinguish multiple samples by color on graph overlays and the zoom and track functions gives the ability to focus on areas of interest.

Results may be downloaded to a USB flash drive in CSV file format or printed to an optional integrated graphical printer for a permanent record. Alternatively, results can be exported via a cable connection to a suitable PC running Print Via Computer (PVC) software or wireless accessory (optional).

 

Comprehensive IQ/OQ Support for your Biowave instrument:


Comprehensive IQ/OQ documentation and service support is available for our Biowave instruments allowing the qualification of wavelength accuracy, wavelength repeatability, photometric accuracy, stray light, resolution, noise, and drift.

If your lab requires IQ (Installation Qualification) and OQ (Operational Qualification), contact us to learn more.

Product name
Item #
 
Technical Specifications
Specification Data
Range 190 to 1100 nm
Monochromator Flat grating
Calibration Automatic upon switch on
Beam Height 15 mm
Spectral Bandwidth 3 nm
Accuracy ±2 nm
Reproducibility ±1 nm
Light Source Xenon flash lamp
Detector Twin CMOS array
Photometric Range -0.300 to 2.500 A, 0.3 to 100 %T
Photometric Linearity ±1.3 % or ±0.008 A whichever is greater at 546 nm
Photometric Reproducibility ±0.002 A to 0.5 A at 546 nm
Stray Light <0.5 %T 340 nm
Stability ±0.01 A/h at 340 nm
Noise ±0.005 peak to peak ±0.002 RMS
Digital Output USB flash drive, PC via PVC software, Optional Wireless
Data Export USB flash drive: .tsv, native PVC format PC via PVC: .csv, .emf, .xlsx, .xls, .rtf, .tsv, native PVC format
Method Storage 90 with PIN number protection
Graphical Display Color touch screen, zoom and track function
Sample ID Yes
Languages English, German, French, Spanish, Italian, Japanese, Chinese
Dimensions 260 × 390 × 100 mm
Weight 3.00 kg 3.00 kg

 

Product name
Item #
 
Related Documents
Biochrom Biowave 3+ CT Color Touch Brochure.pdf DOWNLOAD BROCHURE

 

Product name
Item #
Biowave 3+ CT
Item #:
80-3007-57
Biowave 3+ CT with Printer
Item #:
80-3007-58
Biowave 3+ CT Wireless
Item #:
80-3007-59
Biowave 3+ CT Wireless with Printer
Item #:
80-3007-60
Product name
Item #
Biowave 3+ CT (touchscreen version) IQ/OQ documentation
Item #:
80-3007-73

Document contains all procedures required to ensure the installation and qualification of the system according to the manufacturer’s specification.